AED and first aid kit inspection requirements

What OSHA requires for first aid kits, why AED checks come from manufacturers and state laws, and how to track pad, battery, and supply expirations.

  • OSHA 1910.151
  • OSHA 1926.50
  • ANSI/ISEA Z308.1
  • FDA

Checked against the sources on . General information, not legal advice.

  1. Weekly (construction)

    First aid kit check

    Who
    Employer / site lead
    Record
    Restock log
  2. Monthly

    AED & first aid check

    Who
    Designated person
    Record
    Inspection log
  3. Before expiration

    Replace pads, batteries, supplies

    Who
    EHS / site lead
    Record
    Date-tracked log
  4. After any use

    Restock, swap pads, check unit

    Who
    EHS / program lead
    Record
    Service log
On this page

Every workplace covered by OSHA must keep adequate first aid supplies readily available. General industry follows 29 CFR 1910.151, and construction follows 29 CFR 1926.50. Automated external defibrillators (AEDs) are different: OSHA does not specifically require workplaces to own an AED or set an AED inspection schedule, though OSHA encourages installing them.

Once you put an AED or a first aid kit on the wall, keeping it ready comes down to two habits: regular readiness checks and tracking expiration dates on electrode pads, batteries, and perishable supplies before they lapse.

Who requires what

First aid kits and AEDs sit under different authorities. Knowing which rule applies keeps your inspection program grounded in what an auditor or regulator will actually look for.

  • General industry first aid kits (OSHA 1910.151(b)): Employers must keep adequate first aid supplies readily available. If no infirmary, clinic, or hospital is in near proximity to treat injured employees, someone on site must be adequately trained to render first aid. Non-mandatory Appendix A points to ANSI/ISEA Z308.1 as a benchmark for basic kit contents.
  • Construction first aid kits (OSHA 1926.50(d)(2)): Supplies must be stored in a weatherproof container with individual sealed packages for each item. The employer must check the contents before sending a kit out on each job and at least weekly on each job to replace expended items.
  • Kit contents and containers (ANSI/ISEA Z308.1): A voluntary consensus standard that defines minimum contents (Class A and Class B) and container ratings (Types I through IV). It also states that kits should be inspected frequently and any expired supply discarded and replaced.
  • AEDs (FDA, manufacturers, and state laws): OSHA standards do not specifically require AEDs or mandate a federal inspection interval. Instead, the FDA regulates AEDs and their pads and batteries as premarket-approved medical devices, the manufacturer sets the maintenance and inspection schedule, and state AED laws often require periodic checks, medical oversight, and local EMS registration to maintain Good Samaritan protections.

Monthly AED readiness checks

Most AEDs run automatic daily, weekly, or monthly self-tests, but a self-test only checks internal electronics. It cannot see a torn pad pouch, a missing razor, or an obstructed cabinet. Manufacturers and state AED programs typically call for a monthly visual readiness check (some manuals ask for weekly checks, so follow the model's user guide).

During each check, confirm:

  • Status indicator: The visual readiness light shows ready (usually a flashing green light, checkmark, or "OK" display) and the unit is not chirping or beeping.
  • Cabinet and access: The wall cabinet is unlocked or breaks open cleanly, the alarm works if equipped, signage is visible, and nothing blocks the path to the unit.
  • Exterior and cables: The case has no cracks or moisture damage, and pre-connected electrode cables are plugged in firmly with no frayed insulation.
  • Electrode pads: Both adult pads and pediatric pads (or a pediatric key/attenuator, if your site keeps them) are sealed, undamaged, and within their expiration dates. Good practice is to keep a spare set of adult pads in the case.
  • Battery: The battery is seated properly, shows no corrosion, and is within its install-by or replace-by date.
  • Ready kit: Nitrile gloves, trauma shears, a pocket CPR mask, a disposable razor, and absorbent wipes are present and sealed.

Tracking AED pad and battery expirations

An AED fails most often not from a broken circuit board, but from dried-out electrode gel or a dead battery. Both consumables expire on different clocks, so tracking only the unit's purchase date is not enough.

  • Electrode pads: The conductive hydrogel dries out over time even inside an unopened foil pouch. Most manufacturer pads carry a 2- to 5-year shelf life printed directly on the package. Once opened or used, pads must be replaced immediately.
  • Batteries: Most non-rechargeable AED lithium batteries last 2 to 5 years in standby depending on the model and temperature exposure. Note the difference between the battery's install-by date (the deadline to put a spare battery into the unit) and its standby life (how long it lasts once installed).
  • FDA-approved accessories: The FDA requires AEDs and necessary accessories such as replacement pads and batteries to hold premarket approval (PMA). Buy replacement pads and batteries directly from the manufacturer or authorized distributors, and check the FDA's approved AED and accessory lists if you inherit an older unit.

First aid kit classes and container types

ANSI/ISEA Z308.1 groups workplace kits by the hazards they cover (Class) and the environment the case can handle (Type). OSHA 1910.151 Appendix A notes that a basic ANSI kit covers small worksites, while larger or higher-risk operations should expand supplies based on their own hazard assessment.

ClassificationIntended environmentKey contents or features
Class ACommon low-risk workplaces (offices, retail, light work)Minimum fill for minor cuts, abrasions, burns, eye irritation, and basic bleeding control
Class BComplex or higher-risk sites (construction, warehouses, manufacturing)Broader assortment and larger quantities than Class A, plus a splint and a tourniquet
Type I and IIIndoor useType I is stationary and wall-mounted; Type II is portable for indoor response
Type III and IVMobile, outdoor, or rough environmentsType III is portable and water-resistant; Type IV is portable, waterproof, and built for heavy impact

For construction jobsites, a Type III or Type IV weatherproof case aligns with OSHA 1926.50(d)(2).

What to check in a first aid kit

In construction, OSHA requires checking kits before each job and at least weekly on site. In general industry, OSHA does not fix an interval, so monthly checks (plus a check after every incident where supplies are used) are the standard practice.

  • Tamper seal and restock level: If the tamper-evident seal is broken, open the kit and verify every required item against the contents label inside the lid.
  • Expiration dates: Check antibiotic ointments, antiseptic wipes, burn gel, eye wash bottles, instant cold packs, and any over-the-counter medications. Sterile dressings and adhesive bandages also lose sterility or stickiness past their expiration dates.
  • Package integrity: Pull any item with a torn wrapper, crushed tube, dried-out wipe, or water-damaged box.
  • Bloodborne pathogen supplies: Confirm nitrile exam gloves, CPR breathing barriers, and biohazard disposal bags are intact and stocked for anyone expected to render aid.
  • Location and access: Make sure the kit is clean, clearly marked, unlocked or quickly accessible, and mounted where workers can reach it without hunting for a key.

When to remove an AED or kit from service

Do not leave a dead AED or an empty first aid cabinet on the wall where someone will grab it in an emergency. Pull the unit or restock it right away.

Remove from service if

  • AED status indicator shows red, an "X", a fault icon, or the unit is chirping or beeping.
  • AED battery is low, expired, corroded, or missing.
  • Electrode pads are expired, opened, torn, or missing.
  • AED case is cracked, exposed to extreme heat or freezing temperatures outside the manufacturer's range, or affected by an active FDA recall.
  • AED has just been used on a patient (replace pads, check battery capacity, download event data per your program protocol, and verify self-test passes before returning to service).
  • First aid kit is missing critical items, contaminated by moisture or dirt, or stocked with expired sterile or medicated supplies.

If you pull an AED for service, post a clear sign on the empty cabinet directing staff to the nearest working unit until the replacement is back on the wall.

Records, state rules, and FDA approvals

Federal OSHA does not mandate a specific written log format for monthly first aid kit checks or AED inspections. Even so, keeping dated inspection records matters for three practical reasons:

  • State AED compliance: Many state laws tie civil liability protection for workplace AED programs to documented periodic inspections, current CPR/AED responder training, and post-use reporting.
  • FDA recall tracking: Recording each AED's manufacturer, model, serial number, and Unique Device Identifier (UDI) lets you check the FDA Medical Device Recall database in minutes when a safety notice goes out.
  • Audit and incident readiness: A dated log showing who checked each cabinet, what was restocked, and when pads and batteries were swapped proves your program runs on a schedule rather than after an injury.

GreenTagWorks

What GreenTagWorks does about this

In GreenTagWorks, each AED and first aid kit has its own record, QR or NFC tag, and inspection schedule so technicians can walk a site in rounds from a phone, offline or online. Due and overdue checks show up on the inspector's phone and by site on the web, and a failed check captures a reason and a photo while marking the item out of service until cleared. Every inspection is signed and time-stamped, ready to export as a signed PDF or share link whenever an auditor asks.

Official sources

This guide summarizes publicly available requirements for general information. It is not legal advice. Standards change and local codes, state plans and manufacturers can ask for more, so check the official text, the manufacturer's instructions and your authority having jurisdiction, or ask a qualified safety professional, before relying on it.

Put every inspection on a schedule

Load your equipment, set how often each item is inspected, and inspect from a phone. GreenTagWorks keeps the signed record.